Connecting the Sample Injection Loops

Instrument Guide

| 95

Connecting the Sample Injection Loops

The NGC chromatography systems support a variety of injection loop types and

sizes, including static loops and dynamic loops. Table

6 on page

77 lists the

supported loops.

Note:

You might have already connected a loop if you completed all of the

steps in the section Plumbing and Priming the NGC Chromatography Systems

in the NGC Chromatography Systems and ChromLab Software Installation

Guide.

To connect a loop to the sample inject valve

Connect one end of the loop to Loop E and the other end to Loop F on the
sample inject valve.

During Load, sample enters through Loop F while air, buffer, or excess sample is

pushed out through Loop E to waste.

During Injection

, buffer enters through Loop E and pushes the sample out to the

column through Loop F.