The Detectors

Instrument Guide

| 209

Replacing the Multi-Wavelength UV/Vis Detector
Lamps

This section explains how to replace the deuterium lamp or tungsten lamp inside the

multi-wavelength UV/Vis detector. The replacement lamp unit includes the lamp,

cable connector, and plug. The lamp is replaced as a unit.

Caution:

It is strongly suggested that you wear gloves when you replace the

lamps. Do not touch the lamp glass with bare hands, as oils from your skin will

degrade the lamp over time.

To replace the multi-wavelength UV/Vis lamps

1.

On the touch screen, select Shut Down on the dropdown menu to exit
ChromLab and shut down the NGC instrument.

2.

Disconnect the tubing lines from the top and bottom of the UV cell.

3.

Disconnect the tubing lines from the conductivity monitor.

4.

Remove the multi-wavelength UV/Vis detector module from the instrument.

See

Replacing or Repositioning Modules on the NGC Instruments on page

227

for information about removing a module.

5.

Using a screwdriver, loosen the screw and remove the door on the left side of
the module.