Priming and purging the systems, Priming and purging the systems on – Bio-Rad ChromLab™ Software User Manual

Page 120

Advertising
background image

3

| Preparing the Instrument

118

| NGC Chromatography Systems and ChromLab Software

Priming and Purging the Systems

This section explains how to prime your system before running the pumps, and

how to purge your system of air bubbles or rapidly exchanging buffer

solutions.

Note:

This section assumes that you primed the pump head wash system

when you installed the system. See

Plumbing All NGC Systems on page

265

for

more information.

Priming and purging tasks are specific to the NGC system configuration. Table

13

lists the tasks that are associated with each system configuration. The tasks are

explained in detail in the sections that follow.

Table 13.

Priming and purging tasks for each NGC system configuration

NGC Quest

NGC Scout

NGC Discover

Prime and purge the
inlet port on system
pump A.

Prime and purge Q1–Q4
on the buffer blending
valve.

Tip:

The inlet ports

on system pumps A
and B are primed
and purged when
the buffer blending
valve is primed.

Prime and purge Q1–Q4 on
the buffer blending valve.

Tip:

The inlet ports on

system pumps A and B
are primed and purged
when the buffer
blending valve is
primed.

Prime and purge the
inlet port on system
pump B.

Prime and purge the buffer
inlet valves A and B.

Prime and purge the sample
pump.