Connecting Tubing to the Columns

Note:

Ensure that the system has been purged of air and that the tubing is

filled with buffer before connecting the columns. Review and perform the tasks

in

Priming and Purging the Systems

beginning on page

118.

To connect tubing to the columns

1.

Connect tubing from one of the following modules to the top of the column:

Sample inject valve column port

Column switching valve inlet port (1–5)

2.

Remove the plug from the bottom of the column.

3.

Connect the tubing from one of the following modules to the bottom of the
column:

UV monitor

Column switching valve outlet port (1–5)