Plumbing ngc systems, General guidelines for tubing sizes – Bio-Rad ChromLab™ Software User Manual

Page 266

Advertising
background image

C

| Plumbing the System

264

| NGC Chromatography Systems and ChromLab Software

Plumbing NGC Systems

Several plumbing tasks are required for all system configurations. Other tasks are

specific to the configuration. This section explains how to plumb each instrument

configuration.

General Guidelines for Tubing Sizes

The NGC systems use three sizes of tubing. Use the information in Table

18 to

select the appropriate tubing.

Note:

When plumbing the system, keep tubing lengths to a minimum. Also, all

fittings should be finger tight. Do not overtighten, or you risk damaging the

connection. Bio-Rad offers tubing clips of different sizes in the accessories kit.

For outlet tubing longer than 6", attach a tubing clip in close proximity to the

fitting to keep tubing in place in the event that it detaches from the fitting.

Table 18.

NGC system tubing sizes

Use

Tubing Dimensions

Tubing
Material

Fittings

Vol/cm

Prepump

1/8" OD x 0.062" ID

(3.2 mm OD x 1.6 mm ID)

Clear

PTFE

Super flangeless
for 1/8" (3.2 mm)
OD tubing

20 μl

Postpump
with F10
pump head

1/16" OD x 0.020" ID

(1.6 mm OD x 0.51 mm ID)

Orange

PEEK

Super flangeless
for 1/16" (1.6 mm)
OD tubing

2 μl

Postpump
with F100
pump head

1/16" OD x 0.030" ID

(1.6 mm OD x 0.76 mm ID)

Green

PEEK

Super flangeless
for 1/16" (1.6 mm)
OD tubing

4.5 μl

Advertising