Preparing the columns – Bio-Rad ChromLab™ Software User Manual

Page 142

Advertising
background image

3

| Preparing the Instrument

140

| NGC Chromatography Systems and ChromLab Software

Preparing the Columns

Before a method can be run, the column must be connected to the instrument.

Plumb the column between the sample inject valve and UV detector or the column

switching valve (if installed). You can mount columns using the column clips

included in the accessories kit. Small columns can be mounted either on the front or

the side of the instrument using the cartridge holders, which are capable of holding

1–10 ml cartridges. Large columns should be mounted either on a separate rack or

on the side of the instrument using the larger column holders, which are capable of

holding columns up to 35 mm in size.

Caution:

Prime the system before connecting a column to the NGC instrument

to avoid exposing the column to air trapped in the lines. Ensure the columns are

offline during the priming process. Read the technical information included with

the columns to ensure that the system does not exceed the maximum

allowable pressure.

Advertising