Liquid flow path of the ngc discover pro system, Liquid flow path of the ngc, Discover pro system – Bio-Rad ChromLab™ Software User Manual

Page 29

Advertising
background image

The NGC Instrument Illustrated

Instrument Guide

| 27

Liquid Flow Path of the NGC

Discover Pro System

At startup, ChromLab displays a default fluidic scheme in the System Control tab

(as well as in the Method Editor tab and on the touch screen). The fluidic scheme

depicts the plumbing flow for the modules. The image that follows displays the

plumbing flow of the modules in the NGC Discover Pro system during a sample run.

The modules in the fluidic scheme are described in detail in the sections that follow.

To view or modify module parameters during a manual run

Click the module to access its dialog box.

Alternatively, select the module on the touch screen to access its dialog box.

Tip:

For more information about using ChromLab, see the

NGC

Chromatography Systems and ChromLab Software User Guide.

L

EGEND

1

System pumps A and B

2

Buffer inlet valves A and B

3

Buffer blending valve

4

Sample pump with pressure sensor

5

Sample inlet valve

6

Manual inject syringe

7

Sample inject valve

8

Waste (not provided)

9

Sample loop

10

Column switching valve

11

Column (not provided)

12

Multi-wavelength UV/Vis detector

13

Conductivity monitor

14

pH valve

15

Outlet valve

16

Fraction collector (separate instrument)

17

Waste (not provided)

Advertising