Connecting External Devices to the NGC Instrument

Instrument Guide

| 107

Connecting External Devices to the SIM

WARNING!

These procedures require you to work with exposed wires. Ensure

that your systems are powered off and unplugged before performing this task.

This section explains how to connect different external devices to the SIM. Before

connecting the device, locate the connector ports that you will use on the back of

the SIM. Notice that the terminals differ for different devices:

To import analog signals to the NGC instrument, connect the external

device to either the SIM 1 or SIM 2 ports. See the section

To connect an

external device to the SIM 1 and SIM 2 ports

that follows.

To export digital signals from the NGC instrument, connect the external

device to the DAC A port. Optionally, to trigger the external device to start

measuring data from the NGC instrument, connect a trigger wire to the

DIGITAL OUT port. See the section

To connect an external device to the

DAC A and DIGITAL OUT ports on page

112

.

Tip:

The SIM ships with its connectors inserted into their ports.

To connect an external device to the SIM 1 and SIM 2 ports

1.

On the back of the SIM, firmly grasp the green SIM 1 or SIM 2 connector and
pull it straight out of its port.

2.

Locate the green flathead screwdriver included in the SIM’s accessory kit.