| Maintaining the Instrument

244

| NGC Chromatography Systems and ChromLab Software

20. Route the display cable through the side and rear cable troughs on the

instrument.

21. Locate the cable trough covers and slide mount cover that you removed in

step

5 on page

239.

22. Slide the cable trough covers into place and secure them to the back of the

instrument with 6-32 x 1¼" Phillips screws.

23. Insert the side mount cover on the side of the instrument from which you

removed the touch screen.

24. Insert the touch screen cable into the touch screen connector port on the back

of the instrument and tighten the thumbscrews.