| The NGC Instrument

46

| NGC Chromatography Systems and ChromLab Software

Inlet Valves

The inlet valves automate the use of multiple buffers and samples during method

development. In ChromLab, you can program the inlet valves to

Switch between buffers during a method run

Automate sample selection and loading by way of the sample pump

Switch between cleaning solutions and buffers during system wash and

preparation

The inlet valve has eight defined inlet ports and one outlet port. The active ports are

identified by green LEDs for the buffer inlet valve and blue LEDs for the sample inlet

valve. In ChromLab you can provide a custom label for each port to easily identify it

during method protocol programming.

Tip:

In Manual mode (accessed from the System Control tab in ChromLab),

you can select a specific port on the buffer inlet valve through its dialog box. To

open the dialog box, touch or click the inlet valve in the fluidic scheme. When

the buffer blending valve is present, touch or click the pump/valve module to

access the positions on the buffer inlet valves. To access the positions on the

sample inlet valves, touch or click the sample pump module in the fluidic

scheme.