Plumbing NGC Systems

Instrument Guide

| 287

b.

(Optional) Do one of the following:

Connect Tube #4 from port 1 to the Common inlet port on the BioFrac

fraction collector diverter valve.

Connect Waste tubing to port 1 and immerse in a waste container.

c.

Connect outlet tubing to the remaining ports as needed.

2.

If your system has two outlet valves:

a.

Connect tubing from the Out port on the conductivity monitor or pH
monitor to the In port on outlet valve 1 (O1).

b.

(Optional) Do one of the following:

Connect Tube #4 from port 1 on O1 to the Common inlet port on the

BioFrac fraction collector diverter valve.

Connect Waste tubing to port 1 on O1 and immerse in a waste

container.

c.

Connect tubing from port 12 on O1 to the In port on outlet valve 2 (O2).

d.

Connect outlet tubing to all remaining outlet ports as needed.

3.

Immerse the outlet tubing in fraction collection containers.