Metrohm 796 Titroprocessor User Manual

Page 103

Advertising
background image

6. Sample data and sample silo

796 Titroprocessor, Instructions for use

97

97

Sample identification

The sample identification names and types of field can be configured with
the [Sample idents] softkey in the second softkey bar. These
definitions are also saved when a silo file is stored and can be reloaded with
the silo contents, see page 95.

Field types:

auto

an entry is automatically interpreted as numerical or as
text.

text

an entry (even numerical) is always accepted as being
text and thus cannot be used in any calculations.

numeric

an entry is always interpreted as being numerical and can
thus be used in calculations. (System variables ID1, ID2,
ID3). In this type of field the input of text is not accepted.

ЪДДД

Sample silo: Sample idents

ДДДДДД¿

³ ³

³ Name Field type ³

³ ДДДДДДДДДДДД ДДДДДДДДДДДД ³

³ Ident1 Sample no. auto ³

³ ³

³ Ident2 Origin Text ³

³ ³

³ Ident3 Analysis numer. ³

³ ³

АДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД QUIT ÄÙ

This produces the following type of sample silo.

³ # Methode Sample no. Origin Analysis Size Unit ³

³ ДДД ДДДДДДДД ДДДДДДДДДДДД ДДДДДДДДДДДД ДДДДДДДДДДДД ДДДДДДДДД ДДДДД ³

³ <> 1 971120/2 B5 3455 100 mL ³

If on the main page of the Titroprocessor 796 the silo is switched on then
these field definitions for the sample identifiers will be adopted there.
Otherwise the definitions from the configuration settings will be used, see
page 37.

6.2.2 Automatic data input and output

The sample silo is controlled by pointers. The first silo column always shows
the current position of the input (>) and output markers (<).

Data acceptance from a balance

The input pointer marks the silo line in which the next sample weight will be
accepted from a balance (via an RS232 interface).

The input pointer is set by placing the field cursor on the required silo line
and pressing the [--> In pointer] softkey, which is found on the
second softkey bar. The second softkey bar is accessed with the [ >>
>> ]

softkey.

The input marker (>) is placed on the selected silo line in this way. If sample
data are to be transmitted from a balance then the data are inserted into the
marked silo line and the marker moves downwards by one position.

Advertising