8. Command Reference

796 Titroprocessor, Instructions for use

145

145

The parameters of the monotonic titration modes coincide with those of the
DET modes, see page131ff.

8.3.2 User adjustments

User adjustments allow fine adjustment of the pre-set titration parameters.
Starting from one of the sets of parameters (slow, medium, fast) go to
'Adjustment' on the parameter page and press the [Custom adjust]
softkey to change the individual parameters in the ‘Custom adjust-
ments

’ dialog window.

ЪДДД

MET_PH* Custom adjustments

ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД¿

³ ³

³ Volume steps 0.1 mL Signal drift 50 mV/min ³

³ Titr. rate max mL/Min Equilibr.time auto s ³

³ ³

АДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД QUIT ÄÙ

Volume steps

Size of the fixed volume steps

0,001…0,1…9,999 mL

The size of the volume steps should be selected care-
fully, as excessively large volume steps may adversely
affect the accuracy of the results. A good guide value
for selecting the volume increment is

1

/

20

of the ex-

pected EP volume.

Titr. rate

Titration rate

0,01…150 mL/Min | max

The titration rate has a direct impact on the dosing rate
of the volume increment. The maximum value will de-
pend on the buret cylinder size.

Buret cylinder max. rate

1 mL

3 mL/min

2 mL

6 mL/min

5 mL

15 mL/min

10 mL

30 mL/min

20 mL

60 mL/min

50 mL

150 mL/min

A value up to 150 mL/min can be entered even for small
buret cylinders. When the method is executed, however,
the dosing rate will automatically be reduced to the
largest possible value. Select '

max' whenever possible,

except with viscous reagent solutions.