12.1 Error messages

796 Titroprocessor, Instructions for use

240

240

DIALOG 14:

incomplete parameters

Not all the parameters required for the action to be performed have been
entered. Check the settings.

DOSING

DOSING 1:

lost contact to dosing drive "XX"

Dosing drive XX is not connected. Switch off the Titroprocessor, connect up
the dosing drive and switch on again. Dosing drives are only recognized
correctly when the instrument is started up.

DOSING 2:

already locked "XX"

Dosing drive XX is currently being used in a method and is not available for
manual operations.

DOSING 4:

dosing drive "XX" not found

Dosing drive XX is not connected. Switch off the Titroprocessor, connect up
the dosing drive and switch on again. Dosing drives are only recognized
correctly when the instrument is started up.

DOSING 5:

dosing drive "XX" overload

Dosing drive XX is overloaded. Check the buret unit.

DOSING 7:

dosing drive "XX" not ready

Dosing drive XX is currently being used for a different task. Wait until it is
free again.

DOSING 8:

check exch. unit on drive "XX"

Dosing drive XX cannot carry out the required operation as an ex-
change/dosing unit has not been attached or is not attached correctly.

DOSING 9:

power limit of drive "XX" reached

Dosing drive XX is overloaded. Check the buret unit.

DOSING 10:

command for "XX" not executable

The selected command is not applicable to XX-type dosing drives.

DOSING 11:

common access failure YYYY "XX"

Dosing drive XX does not answer. Check the connections.

DOSING 12:

Prepare dosing unit(s) !

Before a sample series can be started the dosing drive involved should be
correctly prepared. Press the <MAN CONTROL> key, select the corre-
sponding dosing drive and start the PREP cycle with the [Prepare buret]
softkey.