8. Command Reference

796 Titroprocessor, Instructions for use

141

141

pK/HNP Evaluation

pK value and half neutralization potential

on | off

When the pK/HNP evaluation function is switched on,
the potential and the associated volume are filed in the
HNP1…9 resp. HNV1…9 system variables. These vari-
ables can be used for calculations.
Careful pH calibration is essential for pK and HNP
evaluations.
If initial dosing takes place, it must be smaller than ½
the dosed volume at the corresponding equivalence
point, otherwise the pK or HNP value cannot be evalu-
ated.
Note! pK values <3.5 or >10.5 must on principle be
regarded as unreliable, i.e. inaccurate.

Windows

Individual equivalence point windows

Individual EP windows can only be entered if 'window'
was indicated under ‘Equivalence points’.

EP# from … to[ ]

Limits of the individual EP windows

with DET_PH*

pH -20…20 |min | max

with DET_U*, DET_IPOL*

-2000…2000 mV |min | max

with DET_UPOL*

-200 …200 µA |min | max

Different EP windows can be assigned for the various
equivalence points expected. This allows equivalence
points to be assigned with greater reliability to the given
calculation formulae, and suppresses undesirable EPs.
EP windows must not overlap. They must all lie within
the limits of the general EP window. The [Check
limits]

softkey is used to check the correctness of

the limits.
In any one EP window it is always the first equivalence
point found that is recognized. Other EPs are sup-
pressed. On the results page, the corresponding
equivalence point is marked with a '

+' if there is more

than one EP in the corresponding EP window.

EPC [ ]

Equivalence point criterion for EP recognition

0…5…200

An individual EP criterion can be selected for each EP
window. It must, however, always be greater than the EP
criterion for the general EP window.