7.11 Reports

796 Titroprocessor, Instructions for use

126

126

send

Report output via an RS 232 interface

yes | no

A report can be transferred directly to a personal com-
puter via an RS 232 interface. In the configuration of the
interfaces this particular interface must have been defined
as the communications interface; it must be connected to
the receiving instrument and the data transmission mode
'LIMS'

must be switched on. Take care that the trans-

mission parameters (baud rate, etc.) have been set to the
same values on the receiving instrument as for the 796
Titroprocessor.
For storing determination data in a database the Metrohm
database

VESUV 3 is recommended; this can evaluate

the raw data reports of the 796 Titroprocessor. Details can
be found in the

VESUV 3 instruction manual.

FF

Page feed (form feed) at the printer

yes | no

For formatting the printout of several report blocks a page
feed can be defined after each individual report. A 'Form
Feed' only affects the printer. Please note that the internal
thermal printer will only carry out a form feed when 'En-
able auto FF'

is switched on in the interface configu-

ration.

Note: if reports are to be printed out on a page-oriented external printer
(inkjet printer, laser printer) then in order to print out the final page of the re-
port a page feed must be carried out after the last report. This is done by
switching on 'Final form feed' in the determination report, see page
77.