4.5 Stirrer at measuring interface

796 Titroprocessor, Instructions for use

66

66

[Shift

auto]

rotates the rack automatically. The shortest rotary path is
selected.

4.5 Stirrer at measuring interface

Each measuring group of the Titroprocessor 796 is equipped with a stirrer
connection (12 V) to which various stirrers can be connected. These
include:

728 Magnetic stirrer

2.728.0040

722 Rod stirrer

2.722.0010

727 Ti-Stand with rod stirrer

2.727.0010

727 Ti-Stand with magnetic stirrer 2.727.0100

703 Ti-Stand

2.703.0010

The stirrer connection of a measuring group (see rear view of instrument,
page 4) is switched on as standard. It can be switched on the 'Manual
control page' (

<MAN CONTROL> key) with the [Stirrer on/off]

softkey. If necessary select the first measuring group ('Sensors at
interface A/B'

) with the

<PG DN> and <TAB> keys.

The status of the stirrer connection is always shown at the bottom right-hand
side.

Apol ³

³

³

7.000 Stirrer on ³

1.000 ³

³

ДДДДДДДДДДДДДДДДДДДДДДДД QUIT ДЩ

ДДД¿ ЪДДДДДДДВДДДДДДДВДДДДДДД¿

³ ³ ³Stirrer³Remote ³

³ ³ ³on/off ³lines ³

ДДДЩ АДДДДДДДБДДДДДДДБДДДДДДДЩ

4.6 Remote lines (Input / Output)

The [Remote lines] softkey on the 'Manual control page'
(

<MAN CONTROL> key) opens the dialog window of the remote

connection of the Titroprocessor 796.

ЪДДД

Remote lines: A

ДДДДДД¿

³ ³

³ I/O Status 76543210 ³

³ ДДДДДДДДДДДДДДДДДДДДДДДД ³

³ Input 00000001 ³
³ Output

****0001

³

³ ³

АДДДДДДДДДДДДДДДДДДД QUIT ÄÙ

Here the conditions of the logical input lines can be seen and the
conditions of the output lines can be set.
The output lines can be set individually (

1) or reset (0). Lines which are not

to be altered can be given an '

*'. Further details can be found on pages 200f

and 29f.