7. Titration and measuring modes
796 Titroprocessor, Instructions for use
103
103
pH = pK + log(a
B
/a
A
)
If a
B
= a
A
, then pH = pK.
The
pK value determined is the volume at the half neutralization point, which
can be obtained from the titration curve. Careful calibration of the pH elec-
trode used is a precondition for a correct determination of the pK value.
In non-aqueous solutions the half neutralization potential (
HNP) is often
used instead of the
pK value. The HNP value is evaluated in the same way
as the
pK value.
The various parameters for the DET modes are described in detail
on pages 128 to 142.