7. Titration and measuring modes

796 Titroprocessor, Instructions for use

103

103

pH = pK + log(a

B

/a

A

)

If a

B

= a

A

, then pH = pK.

The

pK value determined is the volume at the half neutralization point, which

can be obtained from the titration curve. Careful calibration of the pH elec-
trode used is a precondition for a correct determination of the pK value.
In non-aqueous solutions the half neutralization potential (

HNP) is often

used instead of the

pK value. The HNP value is evaluated in the same way

as the

pK value.

The various parameters for the DET modes are described in detail

on pages 128 to 142.