12

12. . Troubleshooting

Troubleshooting

796 Titroprocessor, Instructions for use

247

247

MEAS.INP. 4:

meas.input 'X': missing A/D-Converter calib. data

Calibration data for measuring group X have been deleted as a result of ex-
ternal influences. Please contact the service department of your local Me-
trohm agency.

MEAS.INP. 5:

meas.input 'X', missing temp. sensor

No temperature sensor has been connected to measuring input X.

MEAS.INP. 6:

meas.input 'X', unable to start, measuring error

MEAS.INP. 7:

meas.input 'X' mode:XXXX, value read error YYYY

MEAS.INP. 8:

meas.input 'X' not accessible

These are serious hardware faults of measuring group X. Please contact the
service department of your local Metrohm agency.

METHOD

METHOD 1:

missing END command in last line XX

No END command has been defined in the method, although this is a pre-
condition for concluding a determination.

METHOD 2:

line XX: too many DET,MET,..,MEAS,CAL modes

A maximum of only 5 titration or measurement modes are allowed in a
method, as there must be sufficient system memory space available for
carrying out the method.

METHOD 3:

line XX: more than one CAL* in (CAL .. )CAL

Only one CAL* command can be defined in a calibration sequence. Several
calibration sequences in one method are allowed.

METHOD 4:

nested (OMOVE sequence, see line XX

The start sequence (OMOVE of the method has been incorrectly pro-
grammed. A sequence must always be concluded with the corresponding
command, in this case OMOVE), before another sequence can be started.

METHOD 5:

nested (CMOVE sequence, see line XX

The closing sequence (CMOVE of the method has been incorrectly pro-
grammed. A sequence must always be concluded with the corresponding
command, in this case CMOVE), before another sequence can be started.

METHOD 7:

nesting error, see line XX

Non-permissible nests of command sequences have been programmed in
the method. A sequence must always be concluded with the corresponding
command before another sequence can be started.