Metrohm 796 Titroprocessor User Manual

Page 142

Advertising
background image

8.2 DET

796 Titroprocessor, Instructions for use

136

136

It is preferable to select a relative stop volume when the
primary objective is to limit the consumption of reagent.

Fill. rate

Buret fill rate

0.01 mL/min…150 mL/min | max.

The buret is automatically filled when the titration ends. The
maximum fill rate will depend on the size of the buret cylin-
der used.

Buret cylinder

max. fill rate

1 mL

3 mL/min

2 mL

6 mL/min

5 mL

15 mL/min

10 mL

30 mL/min

20 mL

60 mL/min

50 mL

150 mL/min

A value up to 150 mL/min can be entered even for small
buret cylinders. When the method is executed, however,
the dosing rate will automatically be reduced to the largest
possible value. Select '

max' whenever possible, except with

viscous reagent solutions which could cause underpres-
sure in the cylinder if the filling rate is too fast (formation of
bubbles!).

Meas.val

Measured value as a discontinuation criteria

with DET_PH:

pH -20 … 20 | off

with DET_U, DET_IPOL:

-2000 … 2000 mV | off

with DET_UPOL:

-200 … 200 µA | off

The measured value at which the titration is discontinued.

EP#

Number of equivalence points

0…9 | off

The titration is discontinued as soon as the pre-set number
of equivalence points is recognized. In principle, only 9
equivalence points can be evaluated. In the setting
'EP# 0' the titration is discontinued after the start condi-
tions are executed (initial dosing). 'off' means, the num-
ber of equivalence points found is not taken into account
for the discontinuation of a titration.

Stop time

Absolute titration duration

1…9999 s | off

The titration is discontinued on expiry of the pre-set stop
time.