13.5 Validation / GLP

796 Titroprocessor, Instructions for use

284

284

The method must contain the following results from the calculation formula:

Result for the statistical calculations
RT1=…

Theoretical result as numerical value
RT2=…

Endpoint which is to be used for the calculation of the systematic error
RE=EPx

(x=0…9)

Example of a titer determination for hydrochloric acid with TRIS as sample
substance.

ЪДДД * Formula ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД¿

³ ³

³ # Comment Formula (R$$, C##, X$$, EP# ...) Unit Dec Type ³

³ ДД ДДДДДДДДДДД ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД ДДДДДД ДДД ДДДДДД ³

³ * 1 Titer RT1=SS*1000/0.1/121.14/EP1 4 stat. ³

³ * 2 Titer(theo) RT2=1.0000 4 stat. ³

³ * 3 Consumpt. RE=EP1 mL 3 stat. ³

The method must be stored.

13.5.4 Preparing for a validation

Call up the main page and press the

<GLP> key.

ЪДДД GLP ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД¿

³ ³

³ Method Validations Date Interval Result ³

³ ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД ³

³ 1. HCl Validation method ³

³ 2. ³

³ 3. ³

³ 4. ³

³ 5. ³

³ ³

³ ³

³ ³

³ ³

³ Instrument ident. : 01106 ³

³ Last servicing : 1999-11-25 ³

³ done by : B. Miller ³

³ Program version : 5.796.0010 ³

³ ³

³ ³

³ ³

³ ³

³ ³

³ ³

АДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД QUIT ДЩ

ЪДДДДДДДВДДДДДДДВДДДДДДД¿ ЪДДДДДДДВДДДДДДДВДДДДДДД¿ ЪДДДДДДДВДДДДДДДВДДДДДДД¿

³ ³ ³Delete ³ ³Define ³Extend.³Start ³ ³ ³Sample ³Sensors³

³ ³ ³method ³ ³method ³info ³valid. ³ ³ ³silo ³ ³

АДДДДДДДБДДДДДДДБДДДДДДДЩ АДДДДДДДБДДДДДДДБДДДДДДДЩ АДДДДДДДБДДДДДДДБДДДДДДДЩ

Five different validation methods can be defined for different purposes. Se-
lect a line and press the [Define method] softkey.