3. Configuration

796 Titroprocessor, Instructions for use

51

51

3.4.3 685 Dosimat, exchange units

For the 685 Dosimat exchange units the types of buret units are not defined
as tubing dimensions. In this case the rinsing volume and the number of
rinsing cycles are entered .

The same names can be used as for the dosing units of the 700 Dosino.
The 796 Titroprocessor can determine the type of dosing unit itself.

ÚÄÄÄConfiguration ДДДДДДДДДДДДДДДДДДДДДДДДД(Program version 5.796.0012)ДДДДДДДД¿

³ ³
³ DialÚÄÄÄ

Tubings

ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД¿ ³

³ ³ ³ ³

³ LCD ³ Dosimat 685: Exchange units, preparation ³ ³

³ LCD ³ ³ ³

³ ³ # Type Volume Cycles Dosing rate Filling rate ³ ³

³ Fehl³ ДД ДДДДДДДД ДДДДДДДДДД ДДДДДД ДДДДДДДДДДДДД ДДДДДДДДДДДДД ³ ³

³ Edit³ 1 Default 3.5 mL 2 max mL/min max mL/min ³ ³

³ Edit³ 2 Changer 10.0 mL 2 max mL/min max mL/min ³ ³

³ ³ 3 mL mL/min mL/min ³ ³

³ ³ 4 mL mL/min mL/min ³ ³

³ ³ 5 mL mL/min mL/min ³ ³

³ ³ 6 mL mL/min mL/min ³ ³

³ ³ 7 mL mL/min mL/min ³ ³

³ ³ 8 mL mL/min mL/min ³ ³

³ ³ 9 mL mL/min mL/min ³ ³

³ ³ 10 mL mL/min mL/min ³ ³

³ ³ ³ ³

³ ³ ³ ³

³ ³ ³ ³

³ АДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД QUIT ДЩ ³

³ ³

АДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД QUIT ДЩ

ЪДДДДДДДВДДДДДДДВДДДДДДД¿ ЪДДДДДДДВДДДДДДДВДДДДДДД¿ ЪДДДДДДДВДДДДДДДВДДДДДДД¿

³ ³ ³ ³ ³ ³ ³ 700 ³ ³Delete ³Delete ³ Check ³

³ ³ ³ ³ ³ ³ ³Dosino ³ ³line ³ all ³ type ³

АДДДДДДДБДДДДДДДБДДДДДДДЩ АДДДДДДДБДДДДДДДБДДДДДДДЩ АДДДДДДДБДДДДДДДБДДДДДДДЩ

Type

Name of the buret unit. Two types are already predefined.
(Default=standard settings, see above,
Changer

=standard setting for sample changer

applications, see above.)

Volume

Reagent volume for one rinsing cycle.

Cycles

Number of rinsing cycles to be carried out.

Dosing

rate

The maximum value for the dosing speed depends on the
cylinder size of the buret used.

Buret cylinder

Max. speed

1 mL

3 mL/min

2 mL

6 mL/min

5 mL

15 mL/min

10 mL

30 mL/min

20 mL

60 mL/min

50 mL

150 mL/min

For smaller buret cylinders a value of up to 150 mL/min
can also be entered, but when the function is carried out
the dosing speed will automatically be reduced to the