Metrohm 796 Titroprocessor User Manual

Page 241

Advertising
background image

12

12. . Troubleshooting

Troubleshooting

796 Titroprocessor, Instructions for use

235

235

CHANGER 29:

changer 'X' error; Key[Reset]-->MAN CONTROL

There is a sample changer fault. It is necessary to reset the sample
changer. Press the <MAN CONTROL> key and then the softkeys [Sample
changer] and [Reset].

CHANGER 42:

missing type def. for rack no. "X"

A rack has been defined, but the type of rack has not been entered. Enter
the correct rack type (e.g. M12-0 for a 12-position Metrohm rack) for the
corresponding sample rack.

CHANGER 43:

same code for rack no. "X" and no. "Y"

Rack codes must be unambiguous. Alter the code of one of the racks men-
tioned. Note that the code must always coincide with the physical arrange-
ment of the magnet code of the racks.

CONFIG

CONFIG 1:

"standard file name" too long

The standard file name entered is too long for the type of file name. Shorten
the text to 5 characters.

CONFIG 2:

invalid password

Access to the access control page is protected by a password. Enter the
correct password.

CURVE

CURVE 1:

too few measuring points

In order to produce a measuring curve or a calibration curve at least 4
measurements must be recorded.

CURVE 2:

curve display not possible

Data required for the representation of a curve are not present. Make sure
that the measurement time recording function has been switched on if the
curve is to be produced on the time axis: see page 135.

CURVE 3:

missing scale parameters for X-axis

CURVE 4:

missing scale parameters for Y-axis

All the scale parameters have not been entered.

CURVE 5:

incomplete parameters

Not all the parameters required for the production of a curve have been en-
tered.