13. Appendix

796 Titroprocessor, Instructions for use

275

275

No. of samples

Defines the number of samples to be processed.

*

indefinite number of samples

Silo

all samples in the sample silo

Rack

all samples on the sample rack

edit. yes/no

: with 'no' it is not possible to change

the number of samples in the status window.

Statistics

Switches on or off the statistics function.

edit. yes/no

: with 'no' it is not possible to change

the status of the statistics function in the status window.

No. data records

Defines the number (2…20) of determinations to be in-
cluded in the statistics calculations. This setting also de-
termines the size of the working memory.

edit. yes/no

: with 'no' it is not possible to change

the number of data records in the status window.

Sample changer

Address of the sample changer (1 or 2). A blank entry
stands for 'no sample changer'.

edit. yes/no

: with 'no' it is not possible to change

the address of the sample changer in the status window.

Load methods from

Defines the storage location (Internal storage or Data
card) where a method can be loaded from.

Auto save data

Switches on or off the auto save function. Determination
data will be saved for further re-evaluation.

edit. yes/no

: with 'no' it is not possible to change

the status of the auto save function in the status window.

Save data to

Defines the storage location (Internal storage or Data
card) where determination data can be saved.

edit. yes/no

: with 'no' it is not possible to change

the storage location in the status window.

Clear working area

With 'yes' the working memory, the sample silo, the statis-
tics, the run number etc. will be cleared at start-up of the
instrument or while signing on with a key card.

Show method window

With 'yes' the method selection window will be displayed
automatically after start-up of the instrument or sign-on
with a key card.