. installation installation, External bus, Dosing devices power cables – Metrohm 796 Titroprocessor User Manual

Page 31: 4 sample changers

Advertising
background image

22. . Installation

Installation

796 Titroprocessor, Instructions for use

25

25

700

685

Dosimat B2

Dosimat B1

729

Dosimat B3

Dosimat B4

EBus

Address 1

EBus

Address 2

External Bus

EBus

Address 1

717

717

700

685

Dosimat C2

Dosimat C1

729

Dosimat C3

Dosimat C4

EBus

Address 2

700

685

Dosimat A2

Dosimat A1

700

685

Dosimat A4

Dosimat A3

Line distributor

Dosing

devices

Power cables

796

2.6.4 Sample changers

External Bus

B

One or two 717 Sample changers can be connected to the 'external
bus' sockets of the 796 Titroprocessor. Take care that the address
selector disk is set correctly (see above). Please connect up larger
systems according to the above diagram.
Two 717 Sample changers can be connected at the same time;
however, they cannot be used simultaneously in one automatic
method. It is possible to change the active sample changer within a
method (see command '

CHANGER').

Manual operation of a sample changer while the other sample
changer is busy carrying out a method is possible and does not
present any problems.