Metrohm 796 Titroprocessor User Manual

Page 186

Advertising
background image

8.9 CAL

796 Titroprocessor, Instructions for use

180

180

This measured value drift control can be switched off
('off'). However, this is not recommended for correct
calibration.

Meas. input

Sensor connection sockets

(A1, A2, Adiff, B1, B2, Bdiff)

The measuring inputs

A1, A2 and Apol are located on

the rear panel of the instrument in the measuring group
Sensors A. Adiff stands for the measuring amplifier
switch on the measurement inputs

A1, A2 and Ref.

If a second measuring group (

Sensors B) is present

then the addresses

B1, B2 , Bpol and Bdiff can be

similarly used.

Equilibr.time

Max. delay period before measurement acceptance

0.1…9999 s | auto | off

The whole delay period for measurement acceptance
becomes effective when, with switched-on drift control
(see measurement drift above), the preset value for the
drift is not undercut. Otherwise the measurement will be
accepted when measurement drift is reached.
The whole delay period will also be allowed to elapse if
the drift control is switched off.
The equilibrium time can be switched off. In this case
only the measurement drift is decisive for the measure-
ment acceptance.
With the setting 'Equilibr.time auto' the Titro-
processor will use an equilibrium time which is suitable
for the preset measurement drift value. It is calculated
according to the following equation:

Equilibr time in s

drift mV

.

(

)

/ (

/ min)

.

=

+

+

150

0 01

5

For the standard setting 2 mV this results in an equilib-
rium time of 110 s.

Temperature

Temperature setting or measuring mode

-170…25.0…500°C | auto | off

The temperature can be automatically determined dur-
ing a measurement with the aid of a Pt100/Pt1000 sen-
sor or a combined electrode with built-in temperature
sensor. It is required for automatic pH correction.

numerical

manual input

auto

automatic measurement during cali-
bration

off

no temperature correction