Metrohm 796 Titroprocessor User Manual

Page 251

Advertising
background image

12

12. . Troubleshooting

Troubleshooting

796 Titroprocessor, Instructions for use

245

245

MAN_CAL 4:

buffer out of tolerance

The buffer value found is outside the tolerance range (± 30 mV) of the
standard buffer given. Check sensor and buffer solution.

MAN_CAL 5:

same buffer

The same buffer value has been found twice in succession. The buffer solu-
tion has probably not been changed.

MAN_CAL 6:

temperature out of range

During temperature measurement an invalid value has been found. Check
that the temperature sensor is functioning properly.

MAN_CAL 7:

temperature out of tolerance

For a correct calibration all buffer solutions in a series must have approxi-
mately similar temperatures. A temperature has been found which is outside
the tolerated scatter range of 2°C.

MAN_CAL 8:

meas. value overrange

A potential > 2000 mV has been measured. Check the sensor.

MAN_CAL 9:

meas. value underrange

A potential < -2000 mV has been measured. Check the sensor.

MAN_CAL 12:

max. number of buffers reached

The buffer list is full. No further buffers can be measured.

MAN_CAL 13:

check buffer type and list

The newly entered buffer type does not match the corresponding buffer in
the table. Correct the buffer type or delete the whole calibration.

MAN_CAL 14:

missing temperature sensor, interface 'X'

The necessary temperature sensor has not been connected.

MAN_CAL 15:

invalid standard buffer value

The measured buffer value cannot be allocated to any standard buffer of the
given manufacturer or type.