8. Command reference

796 Titroprocessor, Instructions for use

205

205

message. In this case it is advisable to pick up such a message with the
SCAN_RSx command (see below).

Command Parameters Parameters

ДДДДДДДДДДДДД ДДДДДДДДДДДДДДДДДДДДДДДД ДДДДДДДДДДДДДДДДДДДДДДДД

CTRL_RS1

For longer data or command sequences several

CTRL commands can be

programmed one after another; these are always transmitted as data lines
with max. 50 characters.

Parameter

Character string / data

up to 50 characters

Any letters, numbers and special characters can be used.
Make sure that these can also be interpreted by the re-
ceiving instrument.

SCAN_RS1, SCAN_RS2

Interpret incoming data

In a sequence the

SCN_RSx command has the effect of stopping the

method run until the predefined character string (up to 50 characters) has
been received via the serial RS 232 interface. Incoming data are checked
character by character.

Command Parameters Parameters

ДДДДДДДДДДДДД ДДДДДДДДДДДДДДДДДДДДДДДД ДДДДДДДДДДДДДДДДДДДДДДДД

SCAN_RS1

Parameter

Character string

up to 50 characters

Any combination of letters, numbers and special charac-
ters can be selected.

'*' can be used as place holder for

any character sequence. (If

'*' is to be interpreted as an

ASCII character then

'**' must always be used.) A place

holder can be set within a character string. When the first
part of the character string has been correctly recognized
a search is made for the first appearance of the character
which follows the

'*'. The comparison of the second part

of the character string is carried out here.

If a

SCAN_RS command is used in a series of samples it is recommended

that the RS 232 interface used is first re-initialized. This has the effect of
clearing the input buffer of the interface and is carried out by pressing
[Config.] [Interfaces] [Reset RS232 X] on the Titroprocessor main
page.