Appendix, Quit, All inp measure – Metrohm 796 Titroprocessor User Manual

Page 273: The test is quit by pressing the <quit> key twice

Advertising
background image

13. Appendix

796 Titroprocessor, Instructions for use

267

267

9.

Connect Simulator to Titroprocessor accord. to Fig. 4.

column C (see Fig. 1)

Input 1 :

-U% / direct

±

0.4 mV

10.

Connect Simulator to Titroprocessor accord. to Fig. 5.

column C (see Fig. 1)

Input 2 :

-U% / direct

±

0.4 mV

11.

Connect Simulator to Titroprocessor accord. to Fig. 6 and place the cover over the solar
cell.

column C (see Fig. 1)

Input Ipol :

143 mV

±

20 mV

Input Upol :

21 µA

±

3.0 µA

12.

<QUIT>

13.

Connect Simulator to Titroprocessor accord. to Fig. 7.

14.

<All inp measure>

column E (see Fig. 1)

Input 1 :

Pt1000

±

0.3°C

15.

The test is quit by pressing the <QUIT> key twice.

1

Ind A2 Ind A1 Ref

Pol

Pt100/1000

Pt100

Pt1000

-U%/direkt +U/direkt +U/1G

726

Simulator

1

Ind A2 Ind A1 Ref

Pol

Pt100/1000

Pt100

Pt1000

-U%/direkt +U/direkt +U/1G

726

Simulator

1

Ind A2 Ind A1 Ref

Pol

Pt100/1000

Pt100

Pt1000

-U%/direkt +U/direkt +U/1G

726

Simulator

Fig.

Fig. 22

Fig.

Fig. 33

Fig.

Fig. 44

1

Ind A2 Ind A1 Ref

Pol

Pt100/1000

Pt100

Pt1000

-U%/direkt +U/direkt +U/1G

726

Simulator

1

Ind A2 Ind A1 Ref

Pol

Pt100/1000

Pt100

Pt1000

-U%/direkt +U/direkt +U/1G

726

Simulator

1

Ind A2 Ind A1 Ref

Pol

Pt100/1000

Pt100

Pt1000

-U%/direkt +U/direkt +U/1G

726

Simulator

Fig.

Fig. 55

Fig.

Fig. 66

Fig.

Fig. 77

Advertising