7.3 MET Monotonic equivalence point titration

796 Titroprocessor, Instructions for use

106

106

7.3.5 Evaluation of pK and HNP values

From the data of the titration curve of an acid/base titration the acidic or ba-
sic strength of a sample can be determined. For this purpose the

pK value

or the

HNP value can be determined automatically. The following formula

applies:

pH = pK + log(a

B

/a

A

)

If a

B

= a

A

, then pH = pK.

The

pK value determined is the volume at the half neutralization point, which

can be obtained from the titration curve. Careful calibration of the pH elec-
trode used is a precondition for a correct determination of the pK value.
In non-aqueous solutions the half neutralization potential (

HNP) is often

used instead of the

pK value. The HNP value is evaluated in the same way

as the

pK value.

The various parameters for the MET modes are described in detail

on pages 143 to 147.