13.1 Diagnosis

796 Titroprocessor, Instructions for use

268

268

13.1.10 To check Analog Interface B

With this diagnostic step the analog inputs of Interface B can be checked for their function-
ality. For this test the testbox with article number 1.767.0010 is required.

1.

The procedure is exactly the same as described in Section 0, the only difference
being that under point 3 one has to mark the letter B instead of letter A.

13.1.11 To check Dosimat Interface

With this diagnostic step the interfaces for the dosing units can be checked for their func-
tionality.

1.

Prepare instrument for diagnostic test (see chap. 13.1.2).

2.

<Dosimat interf.>

3.

Without switching the instrument off, connect a 700 Dosino to the interface being
tested.

4.

<Activ. triacs>

Ä

Power supply to the Dosino is activated. This process takes a certain time and is not noticed by the tester.

5.

<Select Dosimat>

6.

With the cursor, select the interface to be tested

(e.g. A1)

(e.g. A1)

7.

<ENTER>

8.

<Regist. wr./rd.>

Ä

The test runs automatically. If a defect is found, an error message appears on the display, otherwise the previous

screen page appears.

9.

<Funct. test>

Ä

The test runs automatically. If a defect is found, an error message appears on the display, otherwise the previous

screen page appears.

10.

Check the other dosing interfaces, following exactly the same procedure as de-
scribed in points 3...9.

11.

The test is quit by pressing <QUIT>.