4.3 Sensors

796 Titroprocessor, Instructions for use

60

60

Measuring parameters

Signal drift

Limits for signal drift

0.5…2…999 mV/min | off

The measurement is only accepted if the measurement
change of the sensor falls below the preset value for the
signal drift. If this is delayed, the measurement will
nevertheless be accepted when the preset equilibration
time (see below) has elapsed.
This signal drift control can be switched off ('off').
However, this cannot be recommended for correct
calibration.

Equilibr.time

Max. equilibration time before measurement
acceptance

0.1…9999 s | auto | off

The whole equilibration time for measurement
acceptance is only effective when, with drift control
switched on (see signal drift above), the preset value for
signal drift is not undercut. Otherwise the measurement
is accepted as soon as the preset signal drift is
achieved.
The whole equilibration time will also be allowed to
elapse if the drift control is switched off.
The equilibration time can be switched off. This means
that only the signal drift is decisive for measurement
acceptance.
In the setting ‘Equilibr.time auto' the
Titroprocessor will calculate an equilibration time suited
to the preset value for the signal drift according to the
following equation:

Equilibr time in s

drift

mV

.

(

)

/ (

/ min)

,

=

+

+

150

0 01

5

For the standard setting 2 mV this results in an
equilibration time of 110 s.

Temperature

Temperature setting or measuring mode

-170…25.0…500°C | auto | off

The temperature can be determined during a
measurement with the help of a Pt100/Pt1000 sensor or
a combined electrode with built-in temperature sensor.
It is required for automatic pH correction.

numeric

manual input

auto

automatic measurement during the
calibration

off

no temperature correction