3 mains connection – Metrohm 796 Titroprocessor User Manual

Page 23

Advertising
background image

22. . Installation

Installation

796 Titroprocessor, Instructions for use

17

17

WARNING - Fire Hazard -

with the same type and rating of fuse

For continued protection replace only

50-60Hz 160VA

2A(TH)

115V

230V

230V

115V: 100V...120V 10%
230V: 220V...240V 10%

2.3 Mains connection

2.3.1 Fuses

Two fuses

S1

S1

and

S2

S2

are built into the 796 Titroprocessor as standard;

both are type 2 ATH (2 A, slow-blow, with high switching capacity, Metrohm
order no. U.600.0107).

Make sure that the instrument is never operated with any different type
of fuse as otherwise there is a fire hazard!

To change blown fuses proceed as follows:

Exchange fuse S1 (zero conductor, 2 ATH)

Remove mains cable from mains connector.

Use a screwdriver to turn fuse holder

S1

S1

to the left until it can be pulled

out.

Pull out the fuse holder, remove the blown fuse and replace it with a
spare fuse (2 ATH).

Insert the fuse holder in the instrument, press it down with a screwdriver
and then turn it to the right to fix it in position.

Exchange fuse S2 (phase, 2 ATH)

Remove mains cable from mains connector.

Use a screwdriver to push fuse holder

S2

S2

upwards until it can be pulled

out.

Pull out the fuse holder, remove the blown fuse and replace it with a
spare fuse (2 ATH).

Insert the fuse holder in the instrument and push until it clicks into place.

Fuse holder

Mains voltage selector

Earth socket

Mains switch

Rear view

S1

S2

Mains connection