Metrohm 796 Titroprocessor User Manual
Page 184
8.8 MEAS
796 Titroprocessor, Instructions for use
178
178
Stop val.
Measured value for terminating the meas. series
for MEAS_PH*
pH -20…20 | off
for MEAS_U*, MEAS_IPOL*
-2000…2000 mV | off
for MEAS_UPOL*
-200 …200 µA | off
for MEAS_T*
-170…500°C | off
If a stop value is set this means that the measuring se-
ries will be prematurely terminated as soon as the de-
fined measuring value is exceeded. The direction in
which the measurements alter is irrelevant.
8.8.3 Curve display in MEAS modes
The possible settings for curve display in the
MEAS modes correspond to
those of the
DET modes with the exception of the representation on the vol-
ume axis and the derived curves.
Quantity Y-Axis
Measuring quantities on the Y-axis
M. value | Temp. | MV,Temp | all
As well as the standard representation of measured
value/volume other representations are possible. Selec-
tion of measuring quantities on the Y-axis:
M. value
pH, mV, µA or °C (depends on mode)
Temp.
only for automatic measurements
MV, Temp measured value and temp. combined
all
all representable curves, combined