8.8 MEAS

796 Titroprocessor, Instructions for use

178

178

Stop val.

Measured value for terminating the meas. series

for MEAS_PH*

pH -20…20 | off

for MEAS_U*, MEAS_IPOL*

-2000…2000 mV | off

for MEAS_UPOL*

-200 …200 µA | off

for MEAS_T*

-170…500°C | off

If a stop value is set this means that the measuring se-
ries will be prematurely terminated as soon as the de-
fined measuring value is exceeded. The direction in
which the measurements alter is irrelevant.

8.8.3 Curve display in MEAS modes

The possible settings for curve display in the

MEAS modes correspond to

those of the

DET modes with the exception of the representation on the vol-

ume axis and the derived curves.

Quantity Y-Axis

Measuring quantities on the Y-axis

M. value | Temp. | MV,Temp | all

As well as the standard representation of measured
value/volume other representations are possible. Selec-
tion of measuring quantities on the Y-axis:

M. value

pH, mV, µA or °C (depends on mode)

Temp.

only for automatic measurements

MV, Temp measured value and temp. combined

all

all representable curves, combined