31

2.6.9 Devices on the remote interface

External peripheral devices such as pH meters, pumps, etc. can be con-
trolled via the remote interface (25-pole socket) of the Titroprocessor. Each
717 Sample changer also has a remote socket which can be controlled via
the

'External bus'.

Remote

E

8 lines are available for signal output (Output 0-7). The remote socket of the
717 Sample changer has 14 output lines available (Output 0-13).

For receiving signals (e.g. the "ready" signal from a Metrohm instrument)
there are 8 lines available (Input 0-7).

The pin occupancy of the remote socket:

Output 6
Output 7
Output 4
Output 2
Output 0
Output 8
Output 9
Output 10
Input 1
Input 3
Input 5
Input 7
Output 11

0 Vo lt

+5 Volt

Output 5
Output 3
Output 1

Output 12
Output 13

Input 0
Input 2
Input 4
Input 6

0 Vo lt

The control lines of the 796 Titroprocessor are purely signal leads (TTL-level)
and should not be misused for supplying devices.
Remote connections can be used with specially assembled cables to solve
complex connection problems in automated systems.

Connection of Metrohm instruments is solved in the following manner with
6.2141.020 Standard remote cable

796

Metrohm instrument

³

796

Metrohm instrument

³

Output 0

–––––––

Input 0

³

Input 0

–––––––

Output 0

Output 1

–––––––

Input 1

³

Input 1

–––––––

Output 1

Output 2

–––––––

Input 2

³

Input 2

–––––––

Output 2

Output 3

–––––––

Input 3

³

Input 3

–––––––

Output 3

Output 4

–––––––

Input 4

³

Input 4

–––––––

Output 4

Output 5

–––––––

Input 5

³

Input 5

–––––––

Output 5

Output 6

–––––––

Input 6

³

Input 6

–––––––

Output 6

Output 7

–––––––

Input 7

³

Input 7

–––––––

Output 7

Output 8

–––––––

Pin 6

Output 9

–––––––

Pin 7

Output 10 –––––––

Pin 8

Output 11

–––––––

Pin 13

Output 12

–––––––

Pin 19

Output 13

–––––––

Pin 20

Output lines 8…13 can only be used
with the 717 Sample changer, but are
not used as input lines for Metrohm in-
struments. They are laid 1:1 on Pin

6…8,13,19…20.

+5V

active= low
inactive = high

active = low
inactive = high

Inputs

Outputs