Metrohm 796 Titroprocessor User Manual

Page 194

Advertising
background image

8.10 Sample changer commands

796 Titroprocessor, Instructions for use

188

188

Special beaker

Name of special beaker

8 characters

The special beaker can be specified by its name. The en-
try field contains a selection list in which first the rack and
then the required special beaker can be selected.

to tower

Rack rotation target

1 | 2

The tower in front of which the special beaker is to be po-
sitioned can be freely selected.

LIFT

Positioning lift

The lift with the titration head can be positioned to the nearest millimeter. It
is recommended that predefined lift positions are used and fixed in the con-
figuration for the specific rack.

Command Parameters Parameters

ДДДДДДДДДДДДД ДДДДДДДДДДДДДДДДДДДДДДДД ДДДДДДДДДДДДДДДДДДДДДДДД

LIFT

to tower 1 to position 0 mm

on tower

Address of the lift

1 | 2

In a 2-tower changer the lifts can be selected at will. The
lift address always refer to the active changer.

to position

Target position of the lift

work | shift. | rinse. | rest | spec | 0…325 mm

The user-defined lift positions (work position, shift posi-
tion, rinsing position, special position) can be defined at
will for each rack, depending on the beaker sizes used.

work position

lift position in which e.g. a titration can
be carried out

shift position

lift position in which the rack can be
freely rotated without damaging the
sensors or other items.

rinsing position

lift position for rinsing the titration
head

rest position

start position at the upper end of
the tower (0 mm)

s

pecial position freely selectable lift position

Advertising