3.2 Configuration of sample changers

796 Titroprocessor, Instructions for use

38

38

dimensions and other settings for this function are
fixed. Details can be found on page 46ff.

[>> >>]

switches to the second softkey bar.

[Service

diagn.]

opens the dialog window for instrument diagnosis . It
is possible to test the various instrument
components and functions in order to check the
functional capabilities of the Titroprocessor 796.
Instrument diagnosis should be carried out with
great care as improper handling could result in
important data being deleted. Certain of the
diagnostic functions are reserved for our service
technicians.
Instrument diagnosis is described on page 260ff.
It is recommended that all the files in the internal
memory as well as the configuration settings (see
softkey [File]) are stored on data cards or on a
PC before the diagnosis function is carried out.

[>> >>]

switches to the second softkey bar.

3.2 Configuration of sample changers

In order to guarantee reliable sample changer operation it is important that
the settings of the connected sample changer 717 and the sample racks
used are configured correctly.
By pressing the softkey [Config.] and then the softkey [Sample
changer]

the configuration of the sample changer can be accessed from

the main page.

ÚÄÄÄConfiguration ДДДДДДДДДДДДДДДДДДДДДДДДД(Program version 5.796.0012)ДДДДДДДД¿
³ ЪДДД

Configuration: Sample changer

ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД¿ ³

³ Dial³ ³ ³

³ ³ Rack definitions ³ ³

³ LCD ³ # Name Beakers Special positions Type Code ³ ³

³ LCD ³ Work Shift Rinse Spec. ³ ³

³ ³ ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД ³ ³

³ Fehl³ 1 12*250mL 12 0 mm 0 mm 0 mm 0 mm M12-0 000001 ³ ³

³ Edit³ 2 16*150mL 16 0 mm 0 mm 0 mm 0 mm M16-0 000010 ³ ³

³ Edit³ 3 24*75mL 24 0 mm 0 mm 0 mm 0 mm M24-0 001000 ³ ³

³ ³ 6 12*150mL 12 0 mm 0 mm 0 mm 0 mm M12-0 100000 ³ ³

³ ³ 5 14*8oz 14 0 mm 0 mm 0 mm 0 mm M14-0 000101 ³ ³

³ ³ 4 14*200mL 14 0 mm 0 mm 0 mm 0 mm M14-0 000011 ³ ³

³ ³ 7 19*100mL 19 0 mm 0 mm 0 mm 0 mm M19-0 001001 ³ ³

³ ³ 8 0 mm 0 mm 0 mm 0 mm ³ ³

³ ³ 9 0 mm 0 mm 0 mm 0 mm ³ ³

³ ³ 10 0 mm 0 mm 0 mm 0 mm ³ ³

³ ³ 11 0 mm 0 mm 0 mm 0 mm ³ ³

³ ³ 12 0 mm 0 mm 0 mm 0 mm ³ ³

³ ³ 13 0 mm 0 mm 0 mm 0 mm ³ ³

³ ³ 14 0 mm 0 mm 0 mm 0 mm ³ ³

³ ³ 15 0 mm 0 mm 0 mm 0 mm ³ ³

³ ³ 16 0 mm 0 mm 0 mm 0 mm ³ ³

АДДДДДАДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД QUIT ÄÙIT ДЩ

ЪДДДДДДДВДДДДДДДВДДДДДДД¿ ЪДДДДДДДВДДДДДДДВДДДДДДД¿ ЪДДДДДДДВДДДДДДДВДДДДДДД¿

³Delete ³ Rack ³ ³ ³Special³Changer³Changer³ ³Delete ³Delete ³ Check ³

³type ³ File ³ ³ ³beakers³ 1 ³ 2 ³ ³line ³ all ³ code ³

АДДДДДДДБДДДДДДДБДДДДДДДЩ АДДДДДДДБДДДДДДДБДДДДДДДЩ АДДДДДДДБДДДДДДДБДДДДДДДЩ

Advertising