Metrohm 796 Titroprocessor User Manual

Page 180

Advertising
background image

8.8 MEAS

796 Titroprocessor, Instructions for use

174

174

Equilibr.time

Max. delay period before measurement acceptance

0.1…9999 s | auto | off

The whole delay period for measurement acceptance
becomes effective when, with switched-on drift control
(see measurement drift above), the preset value for the
drift is not undercut. Otherwise the measurement will be
accepted when the measurement drift is reached.
The whole delay period will also be allowed to elapse if
the drift control is switched off.
The equilibrium time can be switched off. In this case
only the measurement drift is decisive for the measure-
ment acceptance.
With the setting 'Equilibr.time auto' the Titro-
processor will use an equilibrium time which is suitable
for the preset measurement drift value. It is calculated
according to the following equation:

Equilibr time in s

drift mV

.

(

)

/ (

/ min)

.

=

+

+

150

0 01

5

For the standard setting 50 mV this results in an equilib-
rium time of 26 s.

For multiple measurements the equilibrium time must
be shorter than the measuring interval as otherwise the
timing pattern for the measuring intervals cannot be fol-
lowed if the measurement drift is too large.

Temperature

Temperature setting or measuring mode

-170…25.0…500°C | auto | initial

The temperature can be automatically determined dur-
ing a titration with the aid of a Pt100/Pt1000 sensor or a
combined electrode with built-in temperature sensor
and can be used for automatic pH correction (if pH is
being measured).
It is possible to record the temperature throughout a ti-
tration and to output it as a temperature curve.

numerical

manual input (e.g. for pH correction)

auto

automatic measurement during the
whole titration (for temperature curve)

initial

one-off measurement at the start of
the titration mode

The temperature can be included in formulas for calcu-
lations (see page 119ff), e.g. for volume corrections.
The different system variables are listed on page 172
and 129.

Please note: start temperature ST and closing tem-
perature

CT are only available with auto.