Metrohm 796 Titroprocessor User Manual
Page 39
22. . Installation
Installation
796 Titroprocessor, Instructions for use
33
33
With a suitable multiple cable (with special wiring) even several instruments
can be controlled simultaneously via the remote lines. The bit patterns for
the
CTRL_RM and SCAN_RM commands can be combined for this pur-
pose.
2.6.10 LIMS connection
An RS 232 interface can be used for connecting the 796 Titroprocessor to a
LIMS system (
Laboratory Information Management System). 6.2125.060
Connection cable (25-pole to 25-pole) or 6.2125.010 is connected to RS
232 interface socket 1 or 2. The serial connection
COM1 or COM2 on the
Personal Computer must be used.
During the configuration of the interfaces (softkeys [Config.]
[Interfaces]
) take care that no other instrument (balance or printer) has
been defined for the same connection. Under 'Communication Int.'
enter the corresponding RS 232 interface (1 or 2). Under 'Commun.
mode'
enter 'LIMS'. The remaining data transmission parameters must
be set the same for both the PC and the Titroprocessor.
Reports cannot just be printed out but can also be sent directly to the LIMS
system.