22. . Installation

Installation

796 Titroprocessor, Instructions for use

33

33

With a suitable multiple cable (with special wiring) even several instruments
can be controlled simultaneously via the remote lines. The bit patterns for
the

CTRL_RM and SCAN_RM commands can be combined for this pur-

pose.

2.6.10 LIMS connection

An RS 232 interface can be used for connecting the 796 Titroprocessor to a
LIMS system (

Laboratory Information Management System). 6.2125.060

Connection cable (25-pole to 25-pole) or 6.2125.010 is connected to RS
232 interface socket 1 or 2. The serial connection

COM1 or COM2 on the

Personal Computer must be used.
During the configuration of the interfaces (softkeys [Config.]
[Interfaces]

) take care that no other instrument (balance or printer) has

been defined for the same connection. Under 'Communication Int.'
enter the corresponding RS 232 interface (1 or 2). Under 'Commun.
mode'

enter 'LIMS'. The remaining data transmission parameters must

be set the same for both the PC and the Titroprocessor.
Reports cannot just be printed out but can also be sent directly to the LIMS
system.