12. . Troubleshooting

Troubleshooting

796 Titroprocessor, Instructions for use

253

253

SCAN+CTRL

SCAN+CTRL 1:

Remote interface X not present

The selected 'Remote' interface has not been installed or is not able to func-
tion.

STATIST

STATIST 9:

determ. running, reset not possible

The statistics cannot be reset while a determination is being carried out, as
the working memory would be completely cleared. This would also affect
the determination being carried out.

STORAGE

STORAGE 1:

file is locked

The file mentioned is protected against deletion or overwriting. The protec-
tion can be removed in the file manager with the function [Un/Lock file].

STORAGE 2:

file not found

A file with the file name mentioned cannot be found in the selected memory
area. Check whether the file name has been entered correctly.

STORAGE 5:

inadmissible destination for access control files

Access control files cannot be loaded into the working memory directly.
They are loaded under [Config.] [Access control] [File].

STORAGE 6:

Inadmissible destination for configuration files

Configuration files cannot be loaded into the working memory directly. They
are loaded under [Config.] [File].

STORAGE 7:

Inadmissible destination for rack files

Rack files cannot be loaded into the working memory directly. They are
loaded under [Config.] [Sample changer] [Rack File].

STORAGE 8:

invalid determination

The determination file is damaged and can no longer be read. Delete it.

STORAGE 9:

switch on the data "Receive mode"

For the required action the data receive mode of the RS232 interface must
be switched on. This can be carried out under [Config.] [Interfaces].