8.5 SEC

796 Titroprocessor, Instructions for use

158

158

tion and to output it as a temperature curve.

numerical

manual input (e.g. for pH correction)

auto

automatic measurement during the
whole titration (for temperature curve)

initial

one-off measurement at the start of the
titration mode

The temperature can be included in formulas for calcula-
tions (see page 119ff), e.g. for volume corrections. The dif-
ferent system variables are listed on page 149 and 129.
Please note: start temperature ST and closing tempera-
ture

CT are only available with auto.

U(pol)

Polarization voltage (only with SEC_UPOL*)

-1270…400…1270 mV

The polarization voltage is the voltage which is applied to a
polarized electrode during amperometric titration. The re-
sultant measuring quantity is µA.

I(pol)

Polarization current (only with SEC_IPOL*)

-127…1…127 µA

The polarization current is the current which is applied to a
polarized electrode during voltametric titration. The resul-
tant measuring quantity is mV.