12. . Troubleshooting

Troubleshooting

796 Titroprocessor, Instructions for use

259

259

Power

on

WARN. DETSTORE:

corrupt memory detected.

continue anyway ?

clear working

memory ?

Working memory

is cleared

Y

N

Main
page

N

Storage initializing

recommended.

Initialize partially?

RAM init

without GLP Data

Y

Total initialisation: restart

and press <space>

System

halted

Corrupted memory

blocks

N

Y

Do you want to

ß

If data still have to be

saved. Then clear the wor-
king memory with

Stati-

stics 'rst'

.

ß

All determinations in the

working memory are de-
leted.

ß

All files in the working

memory and internal me-
mory are deleted. The
GLP files remain unaffec-
ted.

If all queries are answered with

<N> then the instrument must be switched

on again. A complete original initialization can be carried out by keeping the
space bar pressed down as the instrument is switched on. If this is not done
the same error message will again be displayed.

When a memory error has occurred it is essential to clear the working me-
mory. This can be done on the main page by entering

Statistics 'rst'

.

Save your configuration settings, methods and determination data re-

gularly on data cards or a personal computer!