8. Command reference

796 Titroprocessor, Instructions for use

179

179

8.9 CAL

Automated calibration of pH electrodes is possible as well as manual cali-
bration. Individual buffer sets can be defined and stored in a method. The
calibration can be carried out as a simple run in dialog operation with man-
ual change of the buffer solutions. If a sample changer is used the calibra-
tion procedure can be carried out fully automatically. The necessary sample
changer commands can be programmed in detail in a calibration loop.

The

CAL* mode creates the following system variables:

R$$

End result

X$$

Intermediate result (not shown in report)

C$$

Common variables (automatic assignment)

AP

Asymmetry pH value, pH(as)

SL

Slope of the calibrated electrode

IM

Initial measured value at start of calibration (in mV)

CM

Closing measured value at end of calibration (in mV)

IT

Initial temperature at start of calibration in °C

CT

Closing temperature at end of calibration in °C

MD

Mode duration (total measurement period) in s

No live displays are possible in the calibration mode. However, it is possible
to carry out calculations using existing system variables, see page 119ff.

The calibration data produced are entered on the Devices & Manual Control'
page. The corresponding calibration curve is also available.

8.9.1 Simple calibration

Calibration parameters

CAL*

Calibration

Sensor Signal drift 2 mV/min

Meas. input A1 Equilibr.time auto s
Temperature 25.0

°C

Comment

Comments on calibration mode

(50 characters)

The commentary on the mode is used for identifying the
specific measurement within a determination. This is
chiefly used for more complex methods with several ti-
tration or measuring modes.

Signal drift

Limit for measurement drift

0.5…2…999 mV/Min | off

The measured value is only accepted when the variation
in the sensor measurement falls below the given value
for the measurement drift. If this is delayed then the
measurement will nevertheless be accepted when the
preset delay period (see below) has elapsed.