8.11 Dosing commands

796 Titroprocessor, Instructions for use

194

194

ADD_R

Adding a relative volume

The ADD_R command allows a calculated volume of auxiliary reagent to be
added at any time during a method run. The volume is calculated from the
sample weight with the aid of a factor.

Command Parameters Parameters

ДДДДДДДДДДДДД ДДДДДДДДДДДДДДДДДДДДДДДД ДДДДДДДДДДДДДДДДДДДДДДДД

ADD_R

Reagent Dos. drive A1

concentr. Dos. rate max mL/min

Factor 0.0 Fill. rate max mL/min

The parameters of the ADD_R command largely corre-
spond to those of the ADD_A command. Only a calcula-
tion factor is entered instead of the absolute volume.

Factor

Calculation factor

-99999.0…0.0…99999.0

The effective value of the relative reagent volume is cal-
culated from:

Volume in mL = Sample size

∗∗ factor

.

The maximum reagent volume which can be added is
1000 mL. If a larger value is obtained from the calculation
then the reagent volume will be automatically reduced to
1000 mL.

The other parameters correspond to those of the

ADD_A command, see

page 192.

PREP

Preparing a buret unit

The preparation, i.e. the rinsing of a buret unit cannot only be carried out
manually, but can also be programmed in a method, preferably in a start
sequence (OMOVE) which is carried out once. The necessary rinsing volume
is calculated from the dimensions of the dosing and filling tubing of the ex-
change unit or dosing unit used.

Command Parameters Parameters

ДДДДДДДДДДДДД ДДДДДДДДДДДДДДДДДДДДДДДД ДДДДДДДДДДДДДДДДДДДДДДДД

PREP

Tubing Default Dos. drive A1

Tubing

Name of type of buret unit

8 characters

The type of buret unit defines the tubing parameters and
other settings which are necessary for the correct prepa-
ration of the buret unit. The definition of the various types
is carried out in the configuration under 'Tubings'. The
entry field contains a selection list with the different types
of dosing drive.