Metrohm 796 Titroprocessor User Manual

Page 81

Advertising
background image

5. Methods and determinations

796 Titroprocessor, Instructions for use

75

75

[* Param]

Titration and measuring modes have extended sets of
parameters for setting control parameters, etc. These are
accessed via this softkey. Descriptions of each setting
can be found under command references, page 127ff.

[*Formula]

Titration and measuring modes allow different calculation
formulas valid for the particular mode to be used. The
formula editor is described in section 7.10 (Titration and
measuring modes), page 118ff.

[* Report]

Titration and measuring modes can contain various report
blocks which only apply to that particular mode. These
mode reports form individual parts of the full report and
appear between the introductory (start) determination
report and the concluding (end) determination report. A
description of the report editor is given in section

7.11

(Titration and measuring modes), page 123ff.

[Insert

line]

inserts an empty command line in front of the current line.

[Delete

deletes the current command line. The field cursor must

line]

be in the command column.

[ >> >>]

shows the second softkey bar.

[Copy

copies the current command line in into a local

line]

buffer.

[Cut

copies the marked command into a local buffer

line]

and deletes the current command line.

[Paste

takes the copied command from the local buffer

line]

and inserts it in front of the current command line.

[ >> >>]

shows the first softkey bar.

5.2.4 Determination report

Determination report blocks can be defined as parts of the full report; they
can either be produced at the start of a determination or at the end of a
determination. In addition, reports can be defined which are produced at the
end of a determination when the capacity of the statistics counter is
reached.
Determination reports cannot only be printed out but can also be transferred
directly to a PC database such as the Metrohm software

VESUV 3.