Metrohm 796 Titroprocessor User Manual

Page 33

Advertising
background image

22. . Installation

Installation

796 Titroprocessor, Instructions for use

27

27

2.6.6 Connecting a balance

RS 232 Interface 1

RS 232 Interface 2

G

The following balances can be connected to an RS 232
interface of the 796 Titroprocessor:

Balance

Cable

Sartorius MP8, MC1

6.2134.060

Mettler AB, AG, PR

LC-RS25, supplied with balance

Mettler AM, PM

from Mettler: ME 33995: green wire to pin 2, brown to pin
3, white to pin 7, yellow to pin 20 of the 25-pole plug
+ 6.2125.010 25-pole/9-pole adapter

Mettler interface 016

cable supplied with interface 016: red wire to pin 3, white
wire to pin 7 of the 25-pole plug
+ 6.2125.010 25-pole/9-pole adapter

Mettler interface 011 or 012

6.2125.020 + 6.2125.010

Mettler AT

from Mettler: ME 33995: green wire to pin 2, brown to pin
3, white to pin 7, yellow to pin 20 of the 25-pole plug
+ 6.2125.010 25-pole/9-pole adapter

AND types ER-60, 120, 180, 182

types FR-200, 300
types FX-200, 300, 320

with RS 232 interface (OP-03)

6.2125.020 + 6.2125.010

Precisa, balances with RS 232C in-
terface

6.2125.080 + 6.2125.010

The type of balance must be entered in the Titroprocessor on the configura-
tion page under [Interfaces].

Advertising