5 sec – Metrohm 796 Titroprocessor User Manual

Page 161

Advertising
background image

8. Command reference

796 Titroprocessor, Instructions for use

155

155

8.5 SEC

Conditioning to a preset endpoint is implemented as an independent mode.
Its own set of parameters makes the conditioning mode

SEC independent

from

SET modes. In this way conditioning is very versatile.

A

SEC mode does not produce a measuring point list and therefore no

curves can be shown. No facilities for calculations are provided.

If a

SEC mode is used in preparation for a SET titration then care must be

taken that the parameters of both modes do not differ too much from each
other.

Four different

SEC modes are available for rapid conditioning to a preset

endpoint; they differ only in the measured quantity:

SEC_PH*

measured quantity pH

SEC_U*

measured quantity mV

SEC_UPOL*

measured quantity µA (selectable pol.voltage in mV)

SEC_IPOL*

measured quantity mV (selectable pol.current in µA)

The following information applies to all

SEC modes. Any variations are

noted.

SEC_PH*

Conditioning to preset

endpoint

Reagent Sensor

Concentr. Meas. input A1
Dos. drive A1

applies to

SEC_PH*, SEC_U*, SEC_UPOL*, SEC_IPOL*

SEC modes create the following system variables:

COV

Conditioning volume; reagent consumption for conditioning

DRC

Volume drift in conditioned state

$CRF

Conditioning flag; correctly conditioned yes/no (1 or 0)

These system variables can be used to apply an automatic drift volume re-
duction:

Formula:

Drift red.

R1=DRC/60*(ITD+MD)/1000

mL

EP corr.

R2=EP1-R1

mL

Here it is assumed that owing to external influences a volume drift is effec-
tive during the sample titration whose value corresponds to the volume drift
in a conditioned state (variable

DRC). Its duration is equivalent to the period

between conditioning switch-off and sample titration switch-off
(

ITD+MD * ).

Advertising