13. Appendix

796 Titroprocessor, Instructions for use

265

265

13.1.7 To check Remote interface

With this diagnostic step the interface Remote can be checked for its functionality. For this
test a special test plug with article number 3.496.8510 is required.

1.

Prepare instrument for diagnostic test (see chap. 13.1.2).

2.

Without switching off the instrument plug in test plug 3.496.8510 to socket
"Remote".

3.

<Control lines>

Ä

The test runs automatically. If a defect is found, an error message appears on the display, otherwise the previous

screen page appears.

4.

Without switching off the instrument remove test plug 3.496.8510.

13.1.8 To check RS232 interface

With this diagnostic step the two interfaces RS232 can be checked for their functionality.
For this test a special test plug with article number 3.496.8480 is required.

1.

Prepare instrument for diagnostic test (see chap. 13.1.2).

2.

<RS232 interf.>

3.

Without switching off the instrument plug in test plug 3.496.8480 to socket "RS232
Interface 1".

4.

<Loopbck test>

Ä

The test runs automatically. If a defect is found, an error message appears on the display, otherwise the previous

screen page appears.

5.

<Change Ifc.>

6.

Without switching off the instrument plug in test plug 3.496.8480 to socket "RS232
Interface 2".

7.

<Loopback test>

Ä

The test runs automatically. If a defect is found, an error message appears on the display, otherwise the previous

screen page appears.

8.

Remove test plug 3.496.8480.

9.

The test is quit by pressing <QUIT>.