8.11 Dosing commands

796 Titroprocessor, Instructions for use

192

192

8.11 Dosing commands

The dosing commands allow various auxiliary reagents to be added from
different dosing devices during complex method runs.

ADD_A

Adding an absolute volume

The

ADD_A command allows a preset volume of an auxiliary reagent to be

added at any time during a method run.

Command Parameters Parameters

ДДДДДДДДДДДДД ДДДДДДДДДДДДДДДДДДДДДДДД ДДДДДДДДДДДДДДДДДДДДДДДД

ADD_A

Reagent Dos. drive A1

Concentr. Dos. rate max mL/min

Volume 0.0 mL Fill. rate max mL/min

Reagent

Reagent name

13 characters

If a name is entered here the automatic reagent check can
be used.
At the start of the method a check is made as to whether a
different entry for the current occupancy of the addressed
dosing device has been given under 'Devices & Man-
ual Control'

(

<MAN CONTROL> key). If these do not

coincide an error message is displayed.
No entry in the 'Reagent' field (or on the 'Devices & ...

'

page) prevents an automatic reagent check.

Dos. drive

Dosing device address

(A1…A4, B1…B4, C1…C4)

The connection sockets

Dos. A1…Dos. A4 on the rear

panel of the instrument have the addresses

A1…A4.

Further dosing devices connected to the

'External Bus'

with the help of a Dosimat interface 729 receive the ad-
dresses

B1…B4 (Dosimat interface 1, EBus address 1) or

C1…C4 (Dosimat interface 2, EBus address 2).

Concentr.

Concentration of auxiliary reagent

(numerical | 5 characters)

Two fields for the input of the concentration of the reagent
solution and concentration units. These fields are also used
for reagent checking (see above) and are compared with
the entries in the dialog window 'Devices & Manual
Control

' at the start of a method.