12.1 Error messages

796 Titroprocessor, Instructions for use

248

248

METHOD 8:

incomplete (OMOVE sequence, see line XX

The start sequence (OMOVE of the method has not been completely de-
fined. A sequence must always be concluded with the corresponding com-
mand, in this case OMOVE).

METHOD 9:

incomplete (CMOVE sequence, see line XX

The closing sequence (CMOVE of the method has not been completely de-
fined. A sequence must always be concluded with the corresponding com-
mand, in this case CMOVE).

METHOD 10:

incomplete (CAL sequence, see line XX

The calibration sequence (CAL of the current method has not been com-
pletely defined. A sequence must always be concluded with the corre-
sponding command, in this case CAL).

METHOD 11:

incomplete (CASE sequence, see line XX

METHOD 12:

incomplete (CASE sequence, see line XX

The condition sequence (CASE of the current method has not been com-
pletely defined. A sequence must always be concluded with the corre-
sponding command, in this case CASE).

METHOD 13:

incorrect OMOVE / CMOVE structure, see line XX

Non-permissible nests of command sequences have been programmed in
the method. A sequence must always be concluded with the corresponding
command before another sequence can be started.

METHOD 14:

line XX: illegal nesting, check method!

Non-permissible nests of command sequences have been programmed in
the method. A sequence must always be concluded with the corresponding
command before another sequence can be started.

MONITORING

MONITORING 1:

no more samples in silo

There are no further samples in the silo. The silo pointer is located at the
bottom line. Set the '<-- Out pointer' again. Silo 'rst' on the main page or in
the <Status> window sets the silo pointer to the first line.

MONITORING 2:

missing method XXXXXXXX

A method given in the silo cannot be loaded because it is stored neither in
the internal memory nor on the data card.

MONITORING 3:

check method XXXXXXXX

The given method is faulty and cannot be carried out.